CVE-2017-0199 2 Microsoft, Philips 6 Office, Windows 7, Windows Server 2008 and 3 more 2026-04-22 7.8 High
Microsoft Office 2007 SP3, Microsoft Office 2010 SP2, Microsoft Office 2013 SP1, Microsoft Office 2016, Microsoft Windows Vista SP2, Windows Server 2008 SP2, Windows 7 SP1, Windows 8.1 allow remote attackers to execute arbitrary code via a crafted document, aka "Microsoft Office/WordPad Remote Code Execution Vulnerability w/Windows API."
CVE-2017-0143 3 Microsoft, Philips, Siemens 28 Server Message Block, Windows 10 1507, Windows 10 1511 and 25 more 2026-04-22 8.8 High
The SMBv1 server in Microsoft Windows Vista SP2; Windows Server 2008 SP2 and R2 SP1; Windows 7 SP1; Windows 8.1; Windows Server 2012 Gold and R2; Windows RT 8.1; and Windows 10 Gold, 1511, and 1607; and Windows Server 2016 allows remote attackers to execute arbitrary code via crafted packets, aka "Windows SMB Remote Code Execution Vulnerability." This vulnerability is different from those described in CVE-2017-0144, CVE-2017-0145, CVE-2017-0146, and CVE-2017-0148.
CVE-2025-3425 1 Philips 1 Intellispace Portal 2026-04-15 N/A
The IntelliSpace portal application utilizes .NET Remoting for its functionality. The vulnerability arises from the exploitation of port 755 through the deserialization vulnerability. After analyzing the configuration files, we observed that the server had set the TypeFilterLevel to Full which is dangerous as it can potentially lead to remote code execution using deserialization. This issue affects IntelliSpace Portal: 12 and prior.
CVE-2025-3424 1 Philips 1 Intellispace Portal 2026-04-15 N/A
The IntelliSpace portal application utilizes .NET Remoting for its functionality. The vulnerability arises from the exploitation of port 755 through the "Object Marshalling" technique, which allows an attacker to read internal files without any authentication. This is possible by crafting specific .NET Remoting URLs derived from information enumerated in the client-side configuration files. This issue affects IntelliSpace Portal: 12 and prior.
CVE-2025-3426 1 Philips 1 Intellispace Portal 2026-04-15 N/A
We observed that Intellispace Portal binaries doesn’t have any protection mechanisms to prevent reverse engineering. Specifically, the app’s code is not obfuscated, and no measures are in place to protect against decompilation, disassembly, or debugging. As a result, attackers can reverse-engineer the application to gain insights into its internal workings, which can potentially lead to the discovery of sensitive information, business logic flaws, and other vulnerabilities. Utilizing this flaw, the attacker was able to identify the Hardcoded credentials from PortalUsersDatabase.dll, which contains .NET remoting definition. Inside the namespace PortalUsersDatabase, the class Users contains the functions CreateAdmin and CreateService that are used to initialize accounts in the Portal service. Both CreateAdmin and CreateService functions contain a hardcoded encrypted password along with its respective salt that are set with the function SetInitialPasswordAndSalt. This issue affects IntelliSpace Portal: 12 and prior; Advanced Visualization Workspace: 15.
CVE-2021-42744 1 Philips 4 Mri 1.5t, Mri 1.5t Firmware, Mri 3t and 1 more 2026-04-02 5.5 Medium
Philips MRI 1.5T and MRI 3T Version 5.3 through 5.8.1 does not restrict or incorrectly restricts access to a resource from an unauthorized actor.
CVE-2021-26262 1 Philips 4 Mri 1.5t, Mri 1.5t Firmware, Mri 3t and 1 more 2026-04-02 5.5 Medium
Philips MRI 1.5T and MRI 3T Version 5.3 through 5.8.1 does not restrict or incorrectly restricts access to a resource from an unauthorized actor.
CVE-2021-26248 1 Philips 4 Mri 1.5t, Mri 1.5t Firmware, Mri 3t and 1 more 2026-04-02 5.5 Medium
Philips MRI 1.5T and MRI 3T Version 5.3 through 5.8.1 does not restrict or incorrectly restricts access to a resource from an unauthorized actor.
CVE-2023-40704 1 Philips 1 Vue Pacs 2025-08-27 6.8 Medium
The product does not require unique and complex passwords to be created during installation. Using Philips's default password could jeopardize the PACS system if the password was hacked or leaked. An attacker could gain access to the database impacting system availability and data integrity.
CVE-2025-27955 1 Philips 1 Clinical Collaboration Platform 2025-06-13 6.5 Medium
Clinical Collaboration Platform 12.2.1.5 has a weak logout system where the session token remains valid after logout and allows a remote attacker to obtain sensitive information and execute arbitrary code.
CVE-2025-27954 1 Philips 1 Clinical Collaboration Platform 2025-06-13 6.5 Medium
An issue in Clinical Collaboration Platform 12.2.1.5 allows a remote attacker to obtain sensitive information and execute arbitrary code via the usertoken function of default.aspx.
CVE-2025-27953 1 Philips 1 Clinical Collaboration Platform 2025-06-13 6.5 Medium
An issue in Clinical Collaboration Platform 12.2.1.5 allows a remote attacker to obtain sensitive information and execute arbitrary code via the session management component.
CVE-2020-16247 1 Philips 1 Clinical Collaboration Platform 2025-06-04 6.8 Medium
Philips Clinical Collaboration Platform, Versions 12.2.1 and prior, exposes a resource to the wrong control sphere, providing unintended actors with inappropriate access to the resource.
CVE-2020-16241 1 Philips 2 Suresigns Vs4, Suresigns Vs4 Firmware 2025-06-04 6.3 Medium
Philips SureSigns VS4, A.07.107 and prior does not restrict or incorrectly restricts access to a resource from an unauthorized actor.
CVE-2020-16239 1 Philips 2 Suresigns Vs4, Suresigns Vs4 Firmware 2025-06-04 4.9 Medium
When an actor claims to have a given identity, Philips SureSigns VS4, A.07.107 and prior does not prove or insufficiently proves the claim is correct.
CVE-2020-16237 1 Philips 2 Suresigns Vs4, Suresigns Vs4 Firmware 2025-06-04 2.1 Low
Philips SureSigns VS4, A.07.107 and prior receives input or data, but it does not validate or incorrectly validates that the input has the properties required to process the data safely and correctly.
CVE-2020-16200 1 Philips 1 Clinical Collaboration Platform 2025-06-04 6.5 Medium
Philips Clinical Collaboration Platform, Versions 12.2.1 and prior, does not properly control the allocation and maintenance of a limited resource, thereby enabling an attacker to influence the amount of resources consumed, eventually leading to the exhaustion of available resources.
CVE-2020-16198 1 Philips 1 Clinical Collaboration Platform 2025-06-04 5 Medium
When an attacker claims to have a given identity, Philips Clinical Collaboration Platform, Versions 12.2.1 and prior, does not prove or insufficiently proves the claim is correct.
CVE-2020-14525 1 Philips 1 Clinical Collaboration Platform 2025-06-04 3.5 Low
Philips Clinical Collaboration Platform, Versions 12.2.1 and prior, does not neutralize or incorrectly neutralizes user-controllable input before it is placed in output used as a webpage that is served to other users.
CVE-2020-14518 1 Philips 1 Dreammapper 2025-06-04 5.3 Medium
Philips DreamMapper, Version 2.24 and prior. Information written to log files can give guidance to a potential attacker.